Chapter 16 339
GPIB Command Reference
LIMVSTAR
16. GPIB
Command Reference
( F - N )
LIMVSTAR
Syntax LIMVSTAR <numeric>[DEG|RAD|OHM|SIE|H|F|PCT]
LIMVSTAR?
Description In the creation/edit of the limit line table, sets the limit central value of each segment start
point (specified with the
“LIMSTAR” command on page 334). Use this command, together
with the “LIMDSTAR” command on page 325, when specifying the limits using the
central value and the width.
If segment creation/edit has not been started, executing this command automatically starts
the edit of the edit target segment and then sets the limit central value of the start point.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit, starts the edit of the edit target segment, and then
sets the limit central value of the start point.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[System] - LIMIT TEST - EDIT LIMIT LINE - EDIT - MIDDLE VALUE
<numeric>
Description Limit central value of the segment start point
Range -1E9 to 1E9
Initial value When creating segment 1: 0
When creating an additional segment: Set value of the previous
segment
Unit Varies depending on the measurement parameter (refer to the
explanation of unit in
“SCAL” on page 423).