302 Chapter 16
GPIB Command Reference
DMKR
DMKR
Syntax DMKR {ON|FIX|TRAC|OFF}
DMKR?
Description For the active trace (set with the “TRAC” command on page 457), makes the setting of the
Δ marker. In the coupled marker mode (set with the “MKRCOUP” command on page 352),
the setting is applied to both the A and B traces regardless of the active trace setting. If the
marker function is off, executing this command causes an error and the command is
ignored.
Parameters
Query response {ON|FIX|TRAC|OFF}<newline><^END>
Corresponding
key
[Marker] - ΔMODE MENU - {ΔMKR | FIXED ΔMKR | TRACKING ΔMKR | ΔMKR OFF}
Description
ON Turns on the Δ marker mode and places the Δ marker at the current
marker position.
FIX Turns on the Δ marker mode and places the fixed Δ marker at the
current marker position.
TRAC Turns on the Δ marker mode and places the tracking Δ marker at
the current marker position.
OFF (initial value) Turns off the Δ marker mode and clears the Δ marker.

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals