396 Chapter 16
GPIB Command Reference
PAVER
PAVER
Syntax PAVER {ON|OFF|1|0}
PAVER?
Description Enables/disables the point averaging function.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Bw/Avg] - POINT AVG on OFF
PAVERFACT
Syntax PAVERFACT <numeric>
PAVERFACT?
Description Sets the point averaging count when using the point averaging function. To set the point
averaging count of each segment when creating the list sweep table, also use this
command.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Bw/Avg] - POINT AVG FACTOR
Description
ON or 1 Enables the point averaging function.
OFF or 0
(initial value)
Disables the point averaging function.
<numeric>
Description Point averaging count
Range 1 to 256
Initial value 4 (Note that, if you use this command to create a segment, the
initial value for segment 1 is 1, and the initial value of an additional
segment is the set value of the previous segment.)
Resolution 1

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals