Chapter 16 295
GPIB Command Reference
DCV
16. GPIB
Command Reference
( * , A - E )
DCV
Syntax DCV <numeric>[V]
DCV?
Description Sets the dc bias output level when the dc bias output mode is the voltage mode or
constant-voltage mode. To set the dc bias output level of each segment when creating the
list sweep table, also use this command.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Source] - BIAS MENU - VOLTAGE LEVEL
DEFC
Syntax DEFC
Description Returns all display color settings to the factory states. (No query)
Corresponding
key
[Display] - more 1/2 - MODIFY COLOR - DEFAULT COLORS
<numeric>
Description Output voltage value of the dc bias
Range -40 to 40
Initial value 0 (Note that the initial value is the current set value of the dc bias
output voltage when creating segment 1; the set value of the
previous segment, when creating an additional segment.)
Unit V (volt)
Resolution 1E-3

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals