378 Chapter 16
GPIB Command Reference
OUTPDTRC?
OUTPDTRC?
Syntax OUTPDTRC?
Description Reads out the values of all measurement points in a data trace array (refer to “Internal data
arrays” on page 81). (Query only)
Query response {numeric 1},{numeric 2},..,{numeric NOP×2-1},{numeric NOP×2}<newline><^END>
Reads out the readout and subsidiary readout of the measurement parameter value of each
measurement point
as shown below. Where, NOP is the number of points, and n is an
integer between 1 and NOP.
{numeric n×2-1}: Readout of the n-th measurement point. If the measurement parameter
is a scalar value (for other than COMPLEX Z-Y), the measurement
parameter value is read out. If the measurement parameter is a vector
value (for COMPLEX Z-Y), resistance (for trace A) or conductance
(for trace B) is read out.
{numeric n×2}: Subsidiary readout of the n-th measurement point. If the measurement
parameter is a scalar value (for other than COMPLEX Z-Y), 0 is
always read out. If the measurement parameter is a vector value (for
COMPLEX Z-Y), reactance (for trace A) or susceptance (for trace B)
is reads out.
Corresponding
key
No front panel key is available to execute this function.

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals