EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>4294A

Agilent Technologies 4294A Programming Manual

Agilent Technologies 4294A
518 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #420 background imageLoading...
Page #420 background image
420 Chapter 16
GPIB Command Reference
SAVDTRC
SAVDTRC
Syntax SAVDTRC {ON|OFF|1|0}
SAVDTRC?
Description When saving the internal data array (refer to “Internal data arrays” on page 81) into a file
using the “SAVDASC” command on page 416 and “SAVDDAT” command on page 417,
enables/disables the save of the data trace array.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Save] - DATA - SELECT CONTENTS - TRACE DATA ON off
SAVMEM
Syntax SAVMEM {ON|OFF|1|0}
SAVMEM?
Description When saving the internal data array (refer to “Internal data arrays” on page 81) into a file
using the “SAVDASC” command on page 416 and “SAVDDAT” command on page 417,
enables/disables the save of the memory array.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Save] - DATA - SELECT CONTENTS - MEMORY on OFF
Description
ON or 1
(initial value)
Enables the save of the data trace array.
OFF or 0 Disables the save of the data trace array.
Description
ON or 1 Enables the save of the memory array.
OFF or 0
(initial value)
Disables the save of the memory array.

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals