390 Chapter 16
GPIB Command Reference
OUTPMWID?
OUTPMWID?
Syntax OUTPMWID?
Description For the active trace (set with the “TRAC” command on page 457), if the trace bandwidth
analysis function is on (specified to ON with the “WIDT” command on page 461), reads
out the result of the bandwidth search. (Query only)
Query response {numeric 1},{numeric 2},{numeric 3},{numeric 4},{numeric 5},{numeric 6}<newline><^END>
The readouts are as follows:
{numeric 1}: Bandwidth.
{numeric 2}: Center frequency.
{numeric 3}: Quality factor.
{numeric 4}: Measurement parameter value of the maker position.
{numeric 5}: Value of the left cutoff point sweep parameter value minus the center
value of the sweep range.
{numeric 6}: Value of the right cutoff point sweep parameter value minus the center
value of the sweep range.
If the trace bandwidth analysis function is off, the readouts are invalid.
Corresponding
key
No front panel key is available to execute this function.

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals