Chapter 16 293
GPIB Command Reference
DCOMOPEN{G|C}
16. GPIB
Command Reference
( * , A - E )
DCOMOPEN{G|C}
Syntax DCOMOPENG <numeric>[SIE]
DCOMOPENC <numeric>
DCOMOPEN{G|C}?
Description For fixture compensation data measurement (“COM{A|B|C}” command on page 281), sets
the conductance value (“DCOMOPENG” command) and the capacitance value
(“DCOMOPENC” command) of the OPEN standard used to measure OPEN compensation
data.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Cal] - FIXTURE COMPEN - DEFINE VALUE - {OPEN CONDUCT (G) | OPEN CAP (C)}
<numeric>
Description Conductance value or capacitance value
Range
*1
-1E6 to 1E6
Initial value 0
Unit For conductance value: S (siemens)
For capacitance value: fF (femto farad)
Resolution For conductance value: 1E-12
For capacitance value: 1
*1.The unit of range value is S (siemens) or F (farad).

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals