354 Chapter 16
GPIB Command Reference
MKRLIMVSTAR
MKRLIMVSTAR
Syntax MKRLIMVSTAR
Description In the creation/edit of the limit line table, sets the measured value of the marker position as
the limit central value of the segment start point. This command is used, together with the
“MKRLIMSTAR” command on page 353, usually when setting the sweep parameter value
and measured value at the marker position as the sweep parameter value and limit central
value of the segment start point, respectively.
If segment creation/edit has not been started, executing this command automatically starts
the edit of the edit target segment and then sets the limit central value of the start point.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit, starts the edit of the edit target segment, and then
sets the limit central value of the start point.
If the marker function is off, executing this command causes an error and the command is
ignored. (No query)
Corresponding
key
[System] - LIMIT TEST - EDIT LIMIT LINE - EDIT - more 1/3 - more 2/3
- MKR VALUE STAR
MKRLIMVSTOP
Syntax MKRLIMVSTOP
Description In the creation/edit of the limit line table, sets the measured value of the marker position as
the limit central value of the segment stop point. This command is used, together with the
“MKRLIMSTOP” command on page 353, usually when setting the sweep parameter value
and measured value at the marker position as the sweep parameter value and limit central
value of the segment stop point, respectively.
If segment creation/edit has not been started, executing this command automatically starts
the edit of the edit target segment and then sets the limit central value of the stop point.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit, starts the edit of the edit target segment, and then
sets the limit central value of the stop point.
If the marker function is off, executing this command causes an error and the command is
ignored. (No query)
Corresponding
key
[System] - LIMIT TEST - EDIT LIMIT LINE - EDIT - more 1/3 - more 2/3
- MKR VALUE STOP
MKRMANP
Syntax MKRMANP
Description If the manual sweep function is on, sets the sweep parameter value at the marker position
as the measurement point. (No query)
Corresponding
key
[Marker] - MKR MANUAL SWP

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals