Chapter 16 465
GPIB Command Reference
:PROGram[:SELected]:DELete:ALL
16. GPIB
Command Reference
( S - Z , :PROG )
:PROGram[:SELected]:DELete:ALL
Syntax :PROGram[:SELected]:DELete:ALL
Description Deletes all programs on the 4294A Instrument BASIC editor. (No query)
Corresponding
key
No front panel key is available to execute this function.
:PROGram[:SELected]:DELete:[SELected]
Syntax :PROGram[:SELected]:DELete:[SELected]
Description Deletes a program on the 4294A. (No query)
Corresponding
key
No front panel key is available to execute this function.
:PROGram[:SELected]:EXECute
Syntax :PROGram[:SELected]:EXECute <string>
Description Execute the specified program command. This command can be executed only when the
status of the program is PAUSE or STOP. (No query)
Parameters
Corresponding
key
No front panel key is available to execute this function.
:PROGram[:SELected]:MALLocate
Syntax :PROGram[:SELected]:MALLocate {<numeric>|DEFault}
:PROGram[:SELected]:MALLocate?
Description The 4294A provides no practical function.
Corresponding
key
No front panel key is available to execute this function.
<string>
Description Command you want to execute

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals