282 Chapter 16
GPIB Command Reference
COMST{A|B|C}
COMST{A|B|C}
Syntax COMST{A|B|C} {ON|OFF|1|0}
COMST{A|B|C}?
Description Toggles on and off the OPEN/SHORT/LOAD compensation of the fixture compensation
function. The “COMSTA” command is used to set the OPEN compensation; the
“COMSTB” command, the SHORT compensation; the “COMSTC” command, the LOAD
compensation. If no fixture compensation coefficient has been stored (executed with the
“COM{A|B|C}” command on page 281), executing this command to turn on the
compensation causes an error and the command is ignored.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Cal] - FIXTURE COMPEN - {OPEN on OFF | SHORT on OFF | LOAD on OFF}
CONT
Syntax CONT
CONT?
Description Sets the sweep mode to the auto continuous sweep (CONT). In this mode, sweeps are
repeated automatically and continuously.
Query response {1|0}<newline><^END>
Corresponding
key
[Trigger] - CONTINUOUS
COPA
Syntax COPA
Description Aborts printout. (No query)
Corresponding
key
[Copy] - ABORT
Description
ON or 1 Turns on the compensation.
OFF or 0 (Initial value) Turns off the compensation.
Description
1 (Initial value) The auto continuous sweep is selected as the sweep mode.
0 The auto continuous sweep is not selected as the sweep mode.

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals