EasyManuals Logo

Agilent Technologies 4294A Programming Manual

Agilent Technologies 4294A
518 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #439 background imageLoading...
Page #439 background image
Chapter 16 439
GPIB Command Reference
SMKRP{1-7}
16. GPIB
Command Reference
( S - Z , :PROG )
SMKRP{1-7}
Syntax SMKRP{1|2|3|4|5|6|7} <numeric>
SMKRP{1|2|3|4|5|6|7}?
Description For the active trace (set with the “TRAC” command on page 457), moves a sub marker
(marker number 1-7) to the specified measurement point. In the coupled marker mode (set
with the
“MKRCOUP” command on page 352), regardless of the active trace setting,
moves the sub marker on both the A and B traces. If this command is executed as Query, it
reads out the measurement point number at the sub marker position. If the sub marker is
off, executing this command causes an error and the command is ignored.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
<numeric>
Description Number of the measurement point where you want to move the sub
marker (the number assigned in order from the left edge assuming
that the number of the leftmost measurement point is 1)
Range 1 to the number of points
Initial value 1
Resolution 1

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals