Chapter 16 289
GPIB Command Reference
DCALSHOR{R|L}
16. GPIB
Command Reference
( * , A - E )
DCALSHOR{R|L}
Syntax DCALSHORR <numeric>[OHM]
DCALSHORL <numeric>[H]
DCALSHOR{R|L}?
Description For user calibration data measurement (“CAL{A|B|C}” command on page 274), sets the
resistance value (“DCALSHORR” command) and the inductance value (“DCALSHORL”
command) of the SHORT standard of the calibration kit used to measure SHORT
calibration data.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[CAL] - USER CAL -DEFINE VAUE - {SHORT RESIST (R) | SHORT INDUCT (L)}
<numeric>
Description Resistance value or inductance value
Range -1E6 to 1E6
Initial value 0
Unit For resistance value: Ω (ohm)
For inductance value: H (henry)
Resolution 1E-12

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals