Chapter 16 267
GPIB Command Reference
ANARANG
16. GPIB
Command Reference
( * , A - E )
ANARANG
Syntax ANARANG <numeric 1>,<numeric 2>[HZ|MHZ|V|A]
ANARANG?
Description Sets the analysis range when performing waveform analysis using the waveform analysis
command. You specify the analysis range using the lower limit and the upper limit (or, the
upper limit and the lower limit). If the lower/upper limit does not match with any
measurement point sweep parameter values, the analysis range lower limit is set to the
minimum measurement point larger than the specified value and the analysis range upper
limit is set to the maximum measurement point smaller than the specified value.
If the sweep condition setting is changed, the analysis range is set to the entire sweep
range. If the sweep range setting is zero span, the analysis range is always the entire sweep
range (all measurement points).
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric 1},{numeric 2}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
<numeric 1> <numeric 2>
Description
Analysis range lower
limit (or upper limit)
Analysis range upper
limit (or lower limit)
Range Current sweep range Current sweep range
When the sweep
parameter is
frequency
Initial value 40 110E6
Unit Hz Hz
Resolution 1E-3 1E-3
When the sweep
parameter is
OSC
level (
voltage)
Initial value 0 1
Unit V (volt) V (volt)
Resolution 1E-3 1E-3
When the sweep
parameter is
OSC
level (
current)
Initial value 40E-6 20E-3
Unit A (ampere) A (ampere)
Resolution 1E-6 1E-6
When the sweep
parameter is
dc bias
level (
voltage)
Initial value 0 0
Unit V (volt) V (volt)
Resolution 1E-3 1E-3
When the sweep
parameter is
dc bias
level (
current)
Initial value 0 0
Unit A (ampere) A (ampere)
Resolution 20E-6 20E-6

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals