Chapter 16 421
GPIB Command Reference
SAVMTRC
16. GPIB
Command Reference
( S - Z , :PROG )
SAVMTRC
Syntax SAVMTRC {ON|OFF|1|0}
SAVMTRC?
Description When saving the internal data array (refer to “Internal data arrays” on page 81) into a file
using the “SAVDASC” command on page 416 and “SAVDDAT” command on page 417,
enables/disables the save of the memory trace array.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Save] - DATA - SELECT CONTENTS - TRACE MEN ON off
SAVPSTA
Syntax SAVPSTA
Description Saves the current instrument state and internal data arrays (refer to “Internal data arrays”
on page 81) into the built-in mass storage, flash memory, as a file having the
AUTOREC.STA file name. The AUTOREC.STA file is automatically recalled at
power-on. Therefore, if you execute this command before power-off, the instrument state
and internal data arrays before the power-off can be restored at the next power-on. (No
query)
Corresponding
key
[Save] - POWER ON CONFIG
Description
ON or 1
(initial value)
Enables the save of the memory trace array.