332 Chapter 16
GPIB Command Reference
LIMSEDI
LIMSEDI
Syntax LIMSEDI <numeric>
Description In the creation/edit of the limit line table, starts the edit of the segment specified as the edit
target (set with the
“LIMSEGM” command on page 333). Note that, if a segment is
specified, the edit target segment is changed to the specified segment and then the edit is
started.
After the completion of the parameter setting of the segment, execute the “LIMSDON”
command on page 331 to finish the edit work. After executing this command and the
“LIMSADD” command on page 330 (to start segment edit), executing this command,
without executing the “LIMSDON” command, causes an error and the command is
ignored.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit and then starts the edit of the segment. (No query)
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Corresponding
key
[System] - LIMIT TEST - EDIT LIMIT LINE - EDIT
<numeric>
Description Segment number you want to edit
Range 1 to the number of the segments in the limit line table
Resolution 1

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals