336 Chapter 16
GPIB Command Reference
LIMSTOP
LIMSTOP
Syntax LIMSTOP <numeric>[HZ|MHZ|V|A]
LIMSTOP?
Description In the creation/edit of the limit line table, sets the sweep parameter value of each segment
stop point.
If segment creation/edit has not been started, executing this command automatically starts
the edit of the edit target segment and then sets the sweep parameter value of the stop point.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit, starts the edit of the edit target segment, and then
sets the sweep parameter value of the stop point.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[System] - LIMIT TEST - EDIT LIMIT LINE - EDIT - more 1/3 - STOP
<numeric>
Description Sweep parameter value of the segment stop point
Range same as “STOP” command on page 446
Initial value When creating segment 1: same as “STOP” command
When creating an additional segment: Set value of the previous segment
Unit same as “STOP” command
Resolution same as “STOP” command

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals