Chapter 16 399
GPIB Command Reference
PKDLTX
16. GPIB
Command Reference
( O - R )
PKDLTX
Syntax PKDLTX <numeric>[HZ|MHZ|V|A]
PKDLTX?
Description For the active trace (set with the “TRAC” command on page 457), sets the ΔX value (refer
to Figure 7-1 on page 97) parameter to define the peak searched for in the marker search
function.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Search] - PEAK - PEAK DEF MENU - PEAK DEF: ΔX
<numeric>
Description ΔX value
When the sweep
parameter is
frequency
Range 0 to 110E6
Initial value 500E3
Unit Hz
When the sweep
parameter is
OSC level (voltage)
Range 0 to 1
Initial value 5E-3
Unit V (volt)
When the sweep
parameter
OSC level (current)
Range 0 to 20E-3
Initial value 100E-6
Unit A (ampere)
When the sweep
parameter is
dc bias level (voltage)
Range 0 to 80
Initial value 0.4
Unit V (volt)
When the sweep
parameter is
dc bias level (current)
Range 0 to 200E-3
Initial value 1E-3
Unit A (ampere)

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals