414 Chapter 16
GPIB Command Reference
RPLPPS?
RPLPPS?
Syntax RPLPPS?
Description Within the waveform analysis range specified with the “ANARANG” command on page
267, reads out the difference between the maximum peak measurement parameter value
and the minimum negative peak measurement parameter value and the sweep parameter
value of the peaks (if there are several maximum peaks or minimum negative peaks, the
value of the leftmost peak within the analysis range). If the value cannot be obtained
because no peak or negative peak is detected, 0 is read out.
The peak and negative peak used at the execution of this command satisfies the condition
set with the
“THRR” command on page 452. It is not related to the peak definition in the
marker search function. (Query only)
Query response {numeric 1},{numeric 2},{numeric 3}<newline><^END>
Readouts are as follows:
{numeric 1}: Difference between the maximum peak measurement parameter value
and the minimum negative peak measurement parameter value.
{numeric 2}: Sweep parameter value of the maximum peak.
{numeric 3}: Sweep parameter value of the minimum negative peak.
Corresponding
key
No front panel key is available to execute this function.
RPLRHEI?
Syntax RPLRHEI?
Description Within the waveform analysis range specified with the “ANARANG” command on page
267, calculates the right ripple (measurement parameter value difference between the peak
and its right adjacent negative peak), and reads out the maximum value. If the value cannot
be obtained because no peak or negative peak is detected, 0 is read out.
The peak and negative peak used at the execution of this command satisfies the condition
set with the
“THRR” command on page 452. It is not related to the peak definition in the
marker search function. (Query only)
Query response {numeric}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
RSCO
Syntax RSCO
Description Returns the tint (set with the “TINT” command on page 454), brightness (set with the
“CBRI” command on page 276), and chroma (set with the “COLOR” command on page
281) of the display color of the item selected with the “COLO” command on page 280 to
the factory states. (No query)
Corresponding
key
[Display] - more 1/2 - MODIFY COLOR - MODIFY ITEM COLOR - MODIFY COLOR
- RESET COLOR

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals