DCI
DCI
Syntax DCI <numeric>[A]
DCI?
Description Sets the dc bias output level when the dc bias output mode is the current mode or
constant-current mode. To set the dc bias output level of each segment when creating the
list sweep table, also use this command.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Source] - BIAS MENU - CURRENT LEVEL
<numeric>
Description Output current value of dc bias
Range -0.1 to 0.1
Initial value 0 (Note that the initial value is the current set value of the dc bias
output current, when creating segment 1; the set value of the
previous segment, when creating an additional segment.)
Unit A (ampere)
Resolution 20E-6