EasyManua.ls Logo

Agilent Technologies 4294A - Trgs; Uskey

Agilent Technologies 4294A
518 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 16 459
GPIB Command Reference
TRGS
16. GPIB
Command Reference
( S - Z , :PROG )
TRGS
Syntax TRGS {INT|EXT|BUS|MAN}
TRGS?
Description Selects a trigger source.
Parameters
Query response {INT|EXT|BUS|MAN}<newline><^END>
Corresponding
key
[Trigger] - SOURCE [ ] -{FREE RUN | EXTERNAL | HP-IB/LAN | MANUAL}
USKEY
Syntax USKEY
Description Displays user keys (ON KEY LABELS) of Instrument BAISC. (No query)
Corresponding
key
[System] - IBASIC - ON KEY LABELS
Description
INT (initial value) Specifies the internal trigger.
EXT Specifies the external trigger inputted from the EXT TRIGGER
terminal on the rear panel.
BUS Specifies the GPIB/LAN trigger (trigger by executing the “*TRG”
command on page 261).
MAN Specifies the manual trigger (trigger by the following key sequence
on the front panel:
[Trigger] - SOURCE [ ] - MANUAL).