272 Chapter 16
GPIB Command Reference
BLIGHT
BLIGHT
Syntax BLIGHT {ON|OFF|1|0}
BLIGHT?
Description Toggles on and off the backlight of the LCD screen. If the backlight is off, you cannot read
displayed information on the screen.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
BMON
Syntax BMON {OFF|VOLT|CURR}
BMON?
Description Sets the measurement item in the dc bias level monitor function.
Parameters
Query response {OFF|VOLT|CURR}<newline><^END>
Corresponding
key
[Display] - BIAS LEVEL [ ] - {OFF | VOLT | CURRENT}
Description
ON or 1 (initial value) Turns on the backlight.
OFF or 0 Turns off the backlight.
Description
OFF (initial value) Turns off the dc bias level monitor function.
VOLT Specifies the dc bias voltage level monitor.
CURR Specifies the dc bias current level monitor.

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals