Performance
Parameters
Allows you to specify when the measurement is started.
Immediate: Starts the measurement when you touch the Start button.
Start at: Starts the measurement at the time specified in the Start at
field on the right.
Stop function
Allows you to specify when the measurement ends. Select the relevant
option from the drop-down menu:
Manual stop: Stops the measurement immediately when you touch the
Stop button.
Stop at: Stops the measurement at the time specified in the Stop at
field on the right.
Duration: Performs the measurement for the duration of time specified
on the right.
Memory allocation
Allows you to specify how the measurement will be stored in the Network
Master's memory. Select the relevant option from the drop-down menu:
Use all storage: When Network Master's memory becames full of
measured data, the whole measurement is stopped.
Continuous: When Network Master's memory becames full of
measured data, the oldest records in that memory will be overwritten.
The Network Master's memory size (the file size of measurement results) is
64 Mbytes per one port.
Estimate of test duration
Contains an estimation of the time (in days, hours, minutes and seconds)
before the whole memory will be filled out by the test. During an ongoing
measurement, the estimate will be recalculated periodically.
The titles of ITU-T Recommendation are below.
G.821 Error performance of an international digital connection operating
at a bit rate below the primary rate and forming part of an Integrated
Services Digital Network
G.826 End-to-end error performance parameters and objectives for
international, constant bit-rate digital paths and connections
G.828 Error performance parameters and objectives for international,
constant bit-rate synchronous digital paths
G.829 Error performance events for SDH multiplex and regenerator
sections
G.8201 Error performance parameters and objectives for multi-operator
international paths within the Optical Transport Network (OTN)
M.2100 Performance limits for bringing-into-service and maintenance of
international multi-operator PDH paths and connections
M.2101.1 Performance limits for bringing-into-service and maintenance
of international SDH paths and multiplex sections
M.2401 Error performance limits and procedures for bringing-into-
service and maintenance of multi-operator international paths and
sections within an optical transport network
M.2110 Bringing into service international multi-operator paths, sections
and transmission systems
OTN-related
This item appears if 'BERT on OTN' is running.
OTN
G.8201, M.2401 (M.2110)
Time period
15 minutes, 1 hour, 2 hours, 24 hours, 7 days
Allocation
SDH/SONET/PDH/DSn Applications BERT
MT1000A Network Master Pro Page 162