Description of device parameters Proline Promag 300 Modbus RS485
104 Endress+Hauser
Active level
Navigation Expert → Input → Status input 1 to n → Active level (1351–1 to n)
Description Use this function to determine the input signal level at which the assigned function is
activated.
Selection • High
• Low
Factory setting High
Response time
Navigation Expert → Input → Status input 1 to n → Response time (1354–1 to n)
Description Use this function to enter the minimum time period for which the input signal level must
be present before the selected function is activated.
User entry 5 to 200 ms
Factory setting 50 ms
3.5 "Output" submenu
Navigation Expert → Output
‣
Output
‣
Curr.output 1 to n
→ 104
‣
PFS output 1 to n
→ 118
‣
Relay output 1 to n
→ 138
‣
Double pulse out
→ 145
3.5.1 "Current output 1 to n" submenu
Navigation Expert → Output → Curr.output 1 to n
‣
Curr.output 1 to n
Terminal no. (0379–1 to n)
→ 105