Description of device parameters Proline Promag 300 Modbus RS485
82 Endress+Hauser
ECC clean. cycle (6557)
→ 83
ECC Polarity (6631)
→ 83
ECC
Navigation Expert → Sensor → Process param. → ECC → ECC (6528)
Prerequisite For the following order code:
"Application package", option EC "ECC electrode cleaning"
Description Use this function to enable and disable cyclic electrode cleaning.
Selection • Off
• On
Factory setting Off
ECC duration
Navigation Expert → Sensor → Process param. → ECC → ECC duration (6555)
Prerequisite For the following order code:
"Application package", option EC "ECC electrode cleaning"
Description Use this function to enter the duration of electrode cleaning in seconds.
User entry 0.01 to 30 s
Factory setting 2 s
ECC recov. time
Navigation Expert → Sensor → Process param. → ECC → ECC recov. time (6556)
Prerequisite For the following order code:
"Application package", option EC "ECC electrode cleaning"
Description Use this function to enter the recovery time after electrode cleaning to prevent signal
output interference. The current output values are frozen in the meanwhile.
User entry 1 to 600 s
Factory setting 60 s