Description of device parameters Proline Promag 300 Modbus RS485
84 Endress+Hauser
Ref. density (1892)
→ 86
Temp. source (6712)
→ 84
External temp. (6673)
→ 84
Ref. temperature (1816)
→ 86
Temp. source
Navigation Expert → Sensor → External comp. → Temp. source (6712)
Description Use this function to select the temperature source.
Selection • Off
• Int.temp. sensor
• External value
• Current input 1
*
• Current input 2
*
• Current input 3
*
Factory setting Off
External temp.
Navigation Expert → Sensor → External comp. → External temp. (6673)
Prerequisite The External value option is selected in the Temp. source parameter (→ 84).
Description Use this function to enter the temperature read in from the external device.
User entry Floating point number with sign
Factory setting –273.15 °C
Additional information Dependency
The unit is taken from the Temperature unit parameter (→ 58)