Description of device parameters Proline Promag 300 Modbus RS485
18 Endress+Hauser
Value 1 display
Navigation  Expert → System → Display → Value 1 display (0107)
Prerequisite A local display is provided.
Description Use this function to select one of the measured values to be shown on the local display.
Selection Volume flow
Mass flow
Correct.vol.flow
Flow velocity
Conductivity
*
CorrConductivity
*
Totalizer 1
Totalizer 2
Totalizer 3
Curr.output 1
Curr.output 2
*
Temperature
*
Electronic temp.
Factory setting Volume flow
Additional information Description
If several measured values are displayed at once, the measured value selected here will be
the first value to be displayed. The value is only displayed during normal operation.
The Format display parameter (→  16) is used to specify how many measured
values are displayed simultaneously and how.
Custody transfer mode
Only available for Promag H.
Once the measuring device has been enabled for custody transfer mode, depending on the
custody transfer approval selected the display can switch to show the relevant information.
For detailed information on custody transfer mode, see the Special Documentation for
the device →  7
Dependency
The unit of the displayed measured value is taken from the System units submenu
(→  55).
0% bargraph 1
Navigation  Expert → System → Display → 0% bargraph 1 (0123)
Prerequisite A local display is provided.
* Visibility depends on order options or device settings