Description of device parameters Proline Promag 300 Modbus RS485
128 Endress+Hauser
Val. at max.freq
Navigation  Expert → Output → PFS output 1 to n → Val. at max.freq (0475–1 to n)
Prerequisite One of the following options is selected in the Assign curr. parameter (→  106):
Volume flow
Mass flow
Correct.vol.flow
Flow velocity
Conductivity
*
CorrConductivity
*
Temperature
*
Electronic temp.
Description Use this function to enter the measured value for the end value frequency.
User entry Signed floating-point number
Factory setting Depends on country and nominal diameter
Additional information Description
Use this function to enter the maximum measured value at the maximum frequency. The
selected process variable is output as a proportional frequency.
Dependency
The entry depends on the process variable selected in the Assign freq. parameter
(→  126).
Measuring mode
Navigation  Expert → Output → PFS output 1 to n → Measuring mode (0479–1 to n)
Prerequisite One of the following options is selected in the Assign curr. parameter (→  106):
Volume flow
Mass flow
Correct.vol.flow
Flow velocity
Conductivity
*
CorrConductivity
*
Temperature
*
Electronic temp.
Description Use this function to select the measuring mode for the frequency output.
Selection Forward flow
Forward/Reverse
Rev. flow comp.
Factory setting Forward flow
* Visibility depends on order options or device settings