Proline Promag 300 Modbus RS485 Description of device parameters
Endress+Hauser 91
Additional information Description
Corrected value = (factor × value) + offset
Mass flow offset
Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Mass flow offset (1841)
Description Use this function to enter the zero point shift for the mass flow trim. The mass flow unit
on which the shift is based is kg/s.
User entry Signed floating-point number
Factory setting 0 kg/s
Additional information Description
Corrected value = (factor × value) + offset
Mass flow factor
Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Mass flow factor (1846)
Description Use this function to enter a quantity factor (without time) for the mass flow. This
multiplication factor is applied over the mass flow range.
User entry Positive floating-point number
Factory setting 1
Additional information Description
Corrected value = (factor × value) + offset
Conduct. offset
Navigation  Expert → Sensor → Sensor adjustm. → Variable adjust → Conduct. offset (1848)
Prerequisite The On option is selected in the Conduct. measur. parameter (→  74) parameter.
Description Use this function to enter the zero point shift for the conductivity trim. The conductivity
unit on which the shift is based is S/m.
User entry Signed floating-point number
Factory setting 0 S/m