Description of device parameters Proline Promag 300 Modbus RS485
74 Endress+Hauser
Conduct. damping
Navigation  Expert → Sensor → Process param. → Conduct. damping (1803)
Prerequisite In the Conduct. measur. parameter (→  74), the On option is selected.
Description Use this function to enter a time constant for conductivity damping (PT1 element).
User entry 0 to 999.9 s
Factory setting 0 s
Additional information Description
The damping is performed by a PT1 element
2)
.
User entry
Value = 0: no damping
Value > 0: damping is increased
Damping is switched off if 0 is entered (factory setting).
Temp. damping
Navigation  Expert → Sensor → Process param. → Temp. damping (1886)
Prerequisite One of the following conditions is met:
Order code for "Sensor option", option CI "Medium temperature measurement"
or
The temperature is read into the flowmeter from an external device.
Description Use this function to enter the time constant for temperature damping.
User entry 0 to 999.9 s
Factory setting 0 s
Conduct. measur.
Navigation  Expert → Sensor → Process param. → Conduct. measur. (6514)
Prerequisite The On option is selected in the Conduct. measur. parameter (→  74) parameter.
Description Use this function to enable and disable conductivity measurement.
2) Proportional behavior with first-order lag