Description of device parameters Proline Promag 300 Modbus RS485
114 Endress+Hauser
t
I [mA]
a =
b =
A0028100
I Current
t Time
With Forward/Reverse option
This option is not possible in this case as the values for the 0/4 mA value parameter
(→  108) and 20 mA value parameter (→  109) have different signs.
With Rev. flow comp. option
Flow components outside the span are buffered, balanced and output after a maximum
delay of 60 s.
t
I [mA]
S = A
S
A
A0028101
I Current
t Time
S Flow components saved
A Balancing of saved flow components
Damping out. 1 to n
Navigation  Expert → Output → Curr.output 1 to n → Damping out. 1 to n (0363–1 to n)
Prerequisite One of the following options is selected in the Assign curr. parameter (→  106):
Volume flow
Mass flow
Correct.vol.flow
Flow velocity
Conductivity
*
CorrConductivity
*
Temperature
*
Electronic temp.
One of the following options is selected in the Current span parameter (→  106):
4...20 mA NAMUR
4...20 mA US
4...20 mA
0...20 mA