Description of device parameters Proline Promag 300 Modbus RS485
68 Endress+Hauser
Additional information Result
Example
If the text CAL is entered, the choose list of the Cal. value unit parameter shows the
following options:
• CAL/Nm3
• CAL/m3
• CAL/ft3
• CAL/Sft3
Spec. enth. off.
Navigation Expert → Sensor → System units → User-spec. units → Spec. enth. off. (0584)
Description Use this function to enter the offset for adapting the user-specific calorific value unit
(without volume).
User entry Signed floating-point number
Factory setting 0
Spec. enth. fac.
Navigation Expert → Sensor → System units → User-spec. units → Spec. enth. fac. (0583)
Description Use this function to enter a quantity factor (without volume) for the user-specific calorific
value unit.
User entry Signed floating-point number
Factory setting 1.0
Additional information Example
1 W × min = 60 J → 0.166 W × min = 1 J → user entry: 0.0166
Energy text
Navigation Expert → Sensor → System units → User-spec. units → Energy text (0600)
Description Use this function to enter a text for the user-specific energy unit.
User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)
Factory setting User en.