Proline Promag 300 Modbus RS485 Description of device parameters
Endress+Hauser 203
Flow velocity
Conductivity
*
CorrConductivity
*
Temperature
*
Factory setting Off
Additional information Description
The simulation value of the process variable selected is defined in the Proc. var. value
parameter (→  203).
Proc. var. value
Navigation  Expert → Diagnostics → Simulation → Proc. var. value (1811)
Description Use this function to enter a simulation value for the selected process variable. Subsequent
measured value processing and the signal output use this simulation value. In this way,
users can verify whether the measuring device has been configured correctly.
User entry Depends on the process variable selected
Factory setting 0
Additional information Entry
The unit of the displayed measured value is taken from the System units submenu
(→  55).
Status inp. sim.
Navigation  Expert → Diagnostics → Simulation → Status inp. sim. (1355)
Description Use this function to switch simulation of the status input on and off. The display alternates
between the measured value and a diagnostic message of the "Function check" category (C)
while simulation is in progress.
Selection Off
On
Factory setting Off
* Visibility depends on order options or device settings