Description of device parameters Proline Promag 300 Modbus RS485
204 Endress+Hauser
Additional information Description
The desired simulation value is defined in the Signal level parameter (→  204).
Selection
Off
Simulation for the status input is switched off. The device is in normal measuring mode
or another process variable is being simulated.
On
Simulation for the status input is active.
Signal level
Navigation  Expert → Diagnostics → Simulation → Signal level (1356)
Prerequisite In the Status inp. sim. parameter (→  203), the On option is selected.
Description Use this function to select the signal level for the simulation of the status input. In this
way, users can verify the correct configuration of the status input and the correct function
of upstream feed-in units.
Selection High
Low
Curr.inp 1 to n sim.
Navigation  Expert → Diagnostics → Simulation → Curr.inp 1 to n sim. (1608–1 to n)
Description Option for switching simulation of the current input on and off. The display alternates
between the measured value and a diagnostic message of the "Function check" category (C)
while simulation is in progress.
The desired simulation value is defined in the Value curr.inp 1 to n parameter.
Selection Off
On
Factory setting Off
Additional information Selection
Off
Current simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
On
Current simulation is active.