Proline Promag 300 Modbus RS485 Description of device parameters
Endress+Hauser 33
Diagnostic no. 835 (0702)
→  38
Diagnostic no. 937 (0743)
→  38
Diagnostic no. 938 (0642)
→  38
Diagnostic no. 961 (0736)
→  39
Diagnostic no. 962 (0745)
→  39
Diagnostic no. 043 (Sens.short circ.)
Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 043 (0650)
Description Option for changing the diagnostic behavior of the diagnostic message 043 Sens.short
circ..
Selection Off
Alarm
Warning
Logbook only
Factory setting Warning
Additional information For a detailed description of the options available, see →  32
Diagnostic no. 302 (Verific. active)
Navigation  Expert → System → Diagn. handling → Diagn. behavior → Diagnostic no. 302 (0739)
Description Use this function to change the diagnostic behavior of the diagnostic message 302 Verific.
active.
Selection Alarm
Warning
Factory setting Warning
Additional information For a detailed description of the options available, see →  32