Proline Promag 300 Modbus RS485 Description of device parameters
Endress+Hauser 25
User entry 1 to 10 s
Factory setting 5 s
Additional information Description
This type of alternating display only occurs automatically if the number of measured
values defined exceeds the number of values the selected display format can display
simultaneously.
• The Value 1 display parameter (→ 18) to Value 4 display parameter (→ 23)
are used to specify which measured values are shown on the local display.
• The display format of the displayed measured values is specified using the Format
display parameter (→ 16).
Custody transfer mode
Only available for Promag H.
Once the measuring device has been enabled for custody transfer mode, depending on the
custody transfer approval selected the display can switch between showing the relevant
information and the custody transfer counter.
For detailed information on custody transfer mode, see the Special Documentation for
the device → 7
Display damping
Navigation Expert → System → Display → Display damping (0094)
Prerequisite A local display is provided.
Description Use this function to enter a time constant for the reaction time of the local display to
fluctuations in the measured value caused by process conditions.
User entry 0.0 to 999.9 s
Factory setting 0.0 s
Additional information User entry
Use this function to enter a time constant (PT1 element
1)
) for display damping:
• If a low time constant is entered, the display reacts particularly quickly to fluctuating
measured variables.
• On the other hand, the display reacts more slowly if a high time constant is entered.
Damping is switched off if 0 is entered (factory setting).
1) proportional transmission behavior with first order delay