Description of device parameters Proline Promag 300 Modbus RS485
66 Endress+Hauser
User entry Max. 10 characters such as letters, numbers or special characters (@, %, /)
Factory setting UserCrVol.
Additional information Result
The defined unit is shown as an option in the choose list of the following parameters:
Cor.volflow unit parameter (→  60)
Corr. vol. unit parameter (→  61)
Example
If the text GLAS is entered, the choose list of the Cor.volflow unit parameter (→  60)
shows the following options:
GLAS/s
GLAS/min
GLAS/h
GLAS/d
Corr vol. offset
Navigation  Expert → Sensor → System units → User-spec. units → Corr vol. offset (0602)
Description Use this function to enter the offset for adapting the user-specific corrected volume unit
and corrected volume flow unit (without time).
Value in user-specific unit = (factor × value in base unit) + offset
User entry Signed floating-point number
Factory setting 0
Cor.vol. factor
Navigation  Expert → Sensor → System units → User-spec. units → Cor.vol. factor (0590)
Description Use this function to enter a quantity factor (without time) for the user-specific corrected
volume unit and corrected volume flow unit.
User entry Signed floating-point number
Factory setting 1.0
Density text
Navigation  Expert → Sensor → System units → User-spec. units → Density text (0570)
Description Use this function to enter a text or the user-specific unit of density.