Description of device parameters Proline Promag 300 Modbus RS485
80 Endress+Hauser
Response time
Navigation Expert → Sensor → Process param. → Empty pipe det. → Response time (1859)
Prerequisite In the Empty pipe det. parameter (→ 79), the On option is selected.
Description Use this function to enter the minimum length of time (debouncing time) the signal must
be present for the diagnostic message S862 Pipe empty to be triggered if the measuring
pipe is empty or partially full.
User entry 0 to 100 s
Factory setting 1 s
New adjustment
Navigation Expert → Sensor → Process param. → Empty pipe det. → New adjustment (6560)
Prerequisite The On option is selected in the Empty pipe det. parameter (→ 79).
Description For selecting whether to perform an empty pipe or full pipe adjustment.
Selection • Cancel
• Empty pipe adj.
• Full pipe adjust
Factory setting Cancel
Progress
Navigation Expert → Sensor → Process param. → Empty pipe det. → Progress (6571)
Prerequisite The On option is selected in the Empty pipe det. parameter (→ 79).
Description Use this function to view the progress.
User interface • Ok
• Busy
• Not ok