Description of device parameters Proline Promag 300 Modbus RS485
80 Endress+Hauser
Response time
Navigation  Expert → Sensor → Process param. → Empty pipe det. → Response time (1859)
Prerequisite In the Empty pipe det. parameter (→  79), the On option is selected.
Description Use this function to enter the minimum length of time (debouncing time) the signal must
be present for the diagnostic message S862 Pipe empty to be triggered if the measuring
pipe is empty or partially full.
User entry 0 to 100 s
Factory setting 1 s
New adjustment
Navigation  Expert → Sensor → Process param. → Empty pipe det. → New adjustment (6560)
Prerequisite The On option is selected in the Empty pipe det. parameter (→  79).
Description For selecting whether to perform an empty pipe or full pipe adjustment.
Selection Cancel
Empty pipe adj.
Full pipe adjust
Factory setting Cancel
Progress
Navigation  Expert → Sensor → Process param. → Empty pipe det. → Progress (6571)
Prerequisite The On option is selected in the Empty pipe det. parameter (→  79).
Description Use this function to view the progress.
User interface Ok
Busy
Not ok