Description of device parameters Proline Promag 300 Modbus RS485
92 Endress+Hauser
Additional information Description
Corrected value = (factor × value) + offset
Conduct. factor
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Conduct. factor (1849)
Prerequisite The On option is selected in the Conduct. measur. parameter (→ 74) parameter.
Description Use this function to enter a quantity factor for the conductivity. This multiplication factor
is applied over the conductivity range.
User entry Positive floating-point number
Factory setting 1
Additional information Description
Corrected value = (factor × value) + offset
Corr. vol offset
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Corr. vol offset (1866)
Description Use this function to enter the zero point shift for the corrected volume flow trim. The
corrected volume flow unit on which the shift is based is 1 Nm³/s.
User entry Signed floating-point number
Factory setting 0 Nm³/s
Additional information Description
Corrected value = (factor × value) + offset
Corr. vol factor
Navigation Expert → Sensor → Sensor adjustm. → Variable adjust → Corr. vol factor (1867)
Description Use this function to enter a quantity factor (without time) for the corrected volume flow.
This multiplication factor is applied over the corrected volume flow range.
User entry Positive floating-point number
Factory setting 1