Description of device parameters Proline Promag 300 Modbus RS485
206 Endress+Hauser
Additional information Dependency
The input range is dependent on the option selected in the Current span parameter
(→  106).
FreqOutputSim 1 to n
Navigation  Expert → Diagnostics → Simulation → FreqOutputSim 1 to n (0472–1 to n)
Prerequisite In the Operating mode parameter (→  120), the Frequency option is selected.
Description Use this function to switch simulation of the frequency output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection Off
On
Factory setting Off
Additional information Description
The desired simulation value is defined in the Freq value 1 to n parameter.
Selection
Off
Frequency simulation is switched off. The device is in normal measuring mode or
another process variable is being simulated.
On
Frequency simulation is active.
Freq value 1 to n
Navigation  Expert → Diagnostics → Simulation → Freq value 1 to n (0473–1 to n)
Prerequisite In the FreqOutputSim 1 to n parameter, the On option is selected.
Description Use this function to enter a frequency value for the simulation. In this way, users can verify
the correct adjustment of the frequency output and the correct function of downstream
switching units.
User entry 0.0 to 12 500.0 Hz