Proline Promag 300 Modbus RS485 Country-specific factory settings
Endress+Hauser 213
Nominal diameter
[mm]
(v ~ 2.5 m/s)
[dm³/min]
500 2 000 m
3
/h
600 2 500 m
3
/h
4.1.3 Output current span
Current output 1 to n 4 to 20 mA NAMUR
4.1.4 Pulse value
Nominal diameter
[mm]
( 2 pulse/s)
[dm³]
2 0.005
4 0.025
8 0.1
15 0.2
25 0.5
32 1
40 1.5
50 2.5
65 5
80 5
100 10
125 15
150 0.03 m
3
200 0.05 m
3
250 0.05 m
3
300 0.1 m
3
350 0.1 m
3
400 0.15 m
3
450 0.25 m
3
500 0.25 m
3
600 0.3 m
3
4.1.5 On value low flow cut off
The switch-on point depends on the type of medium and the nominal diameter.
Nominal diameter
[mm]
(v ~ 0.04 m/s)
[m³/h]
2 0.01
4 0.05
8 0.1
15 0.5
25 1