Description of device parameters Proline Promag 300 Modbus RS485
30 Endress+Hauser
Backup state
Navigation Expert → System → Config. backup → Backup state (2759)
Description Displays the status of the data backup process.
User interface • None
• Backup in progr.
• Restore in progr
• Delete in progr.
• Comp. in progr.
• Restoring failed
• Backup failed
Factory setting None
Compar. result
Navigation Expert → System → Config. backup → Compar. result (2760)
Description Displays the last result of the comparison of the data records in the device memory and in
the HistoROM.
User interface • Set. identical
• Set. not ident.
• No backup
• Backup corrupt
• Check not done
• Dataset incomp.
Factory setting Check not done
Additional information Description
The comparison is started via the Compare option in the Config. managem.
parameter (→ 29).
Selection
Options Description
Set. identical The current device configuration of the HistoROM is identical to the backup copy in
the device memory.
If the transmitter configuration of another device has been transmitted to the
device via HistoROM in the Config. managem. parameter, the current device
configuration of the HistoROM is only partially identical to the backup copy in the
device memory: The settings for the transmitter are not identical.
Set. not ident. The current device configuration of the HistoROM is not identical to the backup
copy in the device memory.
No backup There is no backup copy of the device configuration of the HistoROM in the device
memory.
Backup corrupt The current device configuration of the HistoROM is corrupt or not compatible with
the backup copy in the device memory.