Modbus RS485 Register Information Proline Promag 300 Modbus RS485
254 Endress+Hauser
"Modbus data map" submenu
Navigation: Expert → Communication → Modbus data map
Parameter Register Data type Access Selection / User entry / User interface →
Scan list reg.0 to 15 (7114) 0: 5001
1: 5002
2: 5003
3: 5004
4: 5005
5: 5006
6: 5007
7: 5008
8: 5009
9: 5010
10: 5011
11: 5012
12: 5013
13: 5014
14: 5015
15: 5016
Integer Read / Write 1 to 65 535 156
"Web server" submenu
Navigation: Expert → Communication → Web server
Parameter Register Data type Access Selection / User entry / User interface →
Webserv.language (7221) 4219 Integer Read / Write 0 = English
1 = Deutsch
*
2 = Français
*
3 = Español
*
4 = Italiano
*
5 = Nederlands
*
8 = Svenska
*
10 = Bahasa Indonesia
*
11 = 日本語 (Japanese)
*
12 = Portuguesa
*
13 = Polski
*
14 = русский язык(Ru)
*
15 = čeština (Czech)
*
16 = 中文 (Chinese)
*
18 = Türkçe
*
19 = tiếng Việt (Vit)
*
20 = 한국어 (Korean)
*
157
MAC Address (7214) 4210 to 4217.5 String Read Unique 12-digit character string
comprising letters and numbers
157
IP address (7209) 4155 to 4161.5 String Read 4 octet: 0 to 255 (in the particular octet) 158
Subnet mask (7211) 4163 to 4169.5 String Read 4 octet: 0 to 255 (in the particular octet) 158
Default gateway (7210) 4171 to 4177.5 String Read 4 octet: 0 to 255 (in the particular octet) 158
Webserver funct. (7222) 4220 Integer Read / Write 0 = Off
1 = On
159
Login page (7273) 5802 Integer Read / Write 0 = Without header
1 = With header
159
* Visibility depends on order options or device settings