Proline Promag 300 Modbus RS485 Description of device parameters
Endress+Hauser 31
Options Description
Check not done The device configuration of the HistoROM has not yet been compared to the
backup copy in the device memory.
Dataset incomp. The backup copy in the device memory is not compatible with the device.
HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
3.1.3 "Diagn. handling" submenu
Navigation Expert → System → Diagn. handling
‣
Diagn. handling
Alarm delay (0651)
→ 31
‣
Diagn. behavior
→ 32
Alarm delay
Navigation Expert → System → Diagn. handling → Alarm delay (0651)
Description Use this function to enter the time interval until the device generates a diagnostic
message.
The diagnostic message is reset without a time delay.
User entry 0 to 60 s
Factory setting 0 s
Additional information Result
This setting affects the following diagnostic messages:
• 170 coil resistance
• 832 Electronic temp.
• 833 Electronic temp.
• 834 Process temp.
• 835 Process temp.
• 962 empty pipe