Proline Promag 300 Modbus RS485 Description of device parameters
Endress+Hauser 129
Additional information Selection
For a detailed description of the options available, see the Measuring mode
parameter (→  110)
Examples
For a detailed description of the configuration examples, see the Measuring mode
parameter (→  110)
Damping out. 1 to n
Navigation  Expert → Output → PFS output 1 to n → Damping out. 1 to n (0477–1 to n)
Prerequisite One of the following options is selected in the Assign curr. parameter (→  106):
Volume flow
Mass flow
Correct.vol.flow
Flow velocity
Conductivity
*
CorrConductivity
*
Temperature
*
Electronic temp.
Description Use this function to enter a time constant for the reaction time of the output signal to
fluctuations in the measured value.
User entry 0 to 999.9 s
Factory setting 0.0 s
Additional information User entry
Use this function to enter a time constant (PT1 element
4)
) for frequency output damping:
If a low time constant is entered, the current output reacts particularly quickly to
fluctuating measured variables.
On the other hand, the current output reacts more slowly if a high time constant is
entered.
Damping is switched off if 0 is entered (factory setting).
The frequency output is subject to separate damping that is independent of all preceding
time constants.
* Visibility depends on order options or device settings
4) proportional transmission behavior with first order delay