Proline Promag 300 Modbus RS485 Modbus RS485 Register Information
Endress+Hauser 251
Navigation: Expert → Output → PFS output 1 to n
Parameter Register Data type Access Selection / User entry / User interface → 
Switch status 1 to n (0461–1 to n) 1: 2485
2: 2486
3: 9917
Integer Read 1 = Open
6 = Closed
137
Invert outp.sig. (0470–1 to n) 1: 2583
2: 2584
3: 9916
Integer Read / Write 0 = Yes
1 = No
138
* Visibility depends on order options or device settings
"Relay output 1 to n" submenu
Navigation: Expert → Output → Relay output 1 to n
Parameter Register Data type Access Selection / User entry / User interface → 
Terminal no. (0812–1 to n) 1: 8278
2: 8279
3: 8280
Integer Read 0 = Not used
2 = 24-25 (I/O 2)
3 = 22-23 (I/O 3)
139
Relay outp.func. (0804–1 to n) 1: 2488
2: 2489
3: 9876
Integer Read / Write 1 = Open
2 = Diag. behavior
3 = Fl. direct.check
4 = Limit
5 = Digital Output
6 = Closed
139
Assign dir.check (0808–1 to n) 1: 8251
2: 8252
3: 8253
Integer Read / Write 0 = Off
1 = Volume flow
2 = Mass flow
3 = Correct.vol.flow
140
Assign limit (0807–1 to n) 1: 8248
2: 8249
3: 8250
Integer Read / Write 0 = Off
1 = Volume flow
2 = Mass flow
3 = Correct.vol.flow
4 = Conductivity
*
7 = Temperature
*
9 = CorrConductivity
*
16 = Totalizer 1
17 = Totalizer 2
18 = Totalizer 3
37 = Flow velocity
39 = Electronic temp.
140
Assign diag. beh (0806–1 to n) 1: 8245
2: 8246
3: 8247
Integer Read / Write 0 = Alarm
1 = Warning
2 = Alarm or warning
141
Assign status (0805–1 to n) 1: 8272
2: 8273
3: 8274
Integer Read / Write 0 = Low flow cut off
1 = Partial pipe det
141
Switch-off value (0809–1 to n) 1: 8260 to 8261
2: 8262 to 8263
3: 8264 to 8265
Float Read / Write Signed floating-point number 142
Switch-off delay (0813–1 to n) 1: 8254 to 8255
2: 8256 to 8257
3: 8258 to 8259
Float Read / Write 0.0 to 100.0 s 142
Switch-on value (0810–1 to n) 1: 8233 to 8234
2: 8235 to 8236
3: 8237 to 8238
Float Read / Write Signed floating-point number 142
Switch-on delay (0814–1 to n) 1: 8266 to 8267
2: 8268 to 8269
3: 8270 to 8271
Float Read / Write 0.0 to 100.0 s 143