EasyManua.ls Logo

GE Discovery LS - Figure 6-3: Remove the Table Side Covers

GE Discovery LS
358 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 115
3.) Refer to Figure 6-3. Remove the table top side covers. (Each side cover is held in place with
two quarter-turn captive screws, located on the under the table top.)
4.) Refer to Figure 6-3. Move the interlock jumper to the cover off position.
Figure 6-3: Remove the Table Side Covers
5.) Remove the right table side covers from the elevator cover (scissor section) of the table.
a.) Unfasten the two screws and ground connection from two covers on the right side of the
table and one cover on the left side of the table.
Note: Wait until Section 6.2.5.4, after the pedal assembly is removed, to remove the cover from the left
side of the table.
Move jumper to "cover off" position.
Captive Quarter-Turn Screw

Table of Contents