Figure 6-142: Adjust PET Flag
8. Return to Step 21. Characterize the base for the second time.
9. Run another hardware reset.
10. Set up and run a CT-PET scan to test the new characterization values.
6.2.10 Manual Table Wiring
Refer to Figure 6-143 and Figure 6-144 for manual table wiring diagrams.
Figure 6-143: 2328900-2SCH
Loosen hardware with 2.5 mm Allen wrench.
Adjusting Screw