GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 10 - Electrical Connections Page 273
10.3.1.2 Automated and Extended Table Grounding Connections
• For automated table internal wiring diagrams, see 2338513SCH.
• For extended table internal wiring diagrams, see TBD Glen Schneider.
Note 1 - See Discovery LS Site Preparation Manual, Direction 2317947-100, Section 6.3.7.
Note 2 - See one of the following appropriate manuals:
• LightSpeed 2.X Installation Manual, Direction 2243312-100 (for 4-slice SDAS Systems)
• LightSpeed 3.X Installation Manual, Direction 2296439-100 (for 4 or 8-slice MDAS Systems)
• LightSpeed 4.X Installation Manual, Direction 2341957-100 (for 16-slice MDAS Systems)
Table 10-12: Discovery LS Grounding Cables (Automated Tables)
No Part Number Gauge Description From To
1 Hospital
Supplied
See
Note 1.
Power and Ground Hospital Mains PET PDU and CT PDU
2 46-311772 #10 Power and Ground
(4 cables)
PET PDU 3 cables to PET Gantry
1 cable to PET PDB
3 46-327217P1
2275844?
#1/0 GND CT PDU Grounding Bar
4 See Note 2. GND CT Gantry Power
Panel
Grounding Bar
5 2295967 #2 New GND Connection PET Gantry Left Leg Grounding Bar
6 2275584-2 #2 GND CT Console Grounding Bar
7 2295969 #2 Robotic Cable GND ETC Backplane Grounding Bar
8 2275844 #1 GND CT Gantry Grounding Bar
9 2301800 #1 Robotic Cable GND CT Gantry Table
10 2340388 #14 Ground SB Connector Bracket Grounding Bar
11 2323722 Base Axis Motor M SB Connector Bracket Table Motor