GE HEALTHCARE
DIRECTION 2307224-100, REVISION 6DISCOVERY LS SYSTEM SERVICE MANUAL
Chapter 6 - Table Page 157
4.) Refer to Figure 6-64. Attach the Base CMD Adaptor Cable, P/N 2338086, to J2 of the CMD
Adapter, and tighten the screws. (The Base CMD Adapter Cable, P/N 2338086 was installed
in Section 6.2.6.18.
Figure 6-64: Attach Base CMD Adapter Cable to the J2 of CMD Adapter
6.2.6.22 Attach Flanges (P/N 2320158) to the Secondary Base
Attach the cover flanges to the right and left side of the secondary base at this time to prepare for
installation of the Clutch Motor cables to the left flange. (The rest of the cover brackets will be
installed later.) Locate and remove the following hardware from the upgrade kit:
1.) Use a 3 mm Allen wrench to fasten the cover flange to the right secondary base side rail with
eight (8) sets of M4 hardware. (If necessary, finish tapping the mounting holes on both sides
of the base.)
2.) Use a 3 mm Allen wrench to fasten the cover flange to the left secondary base side rail with
five sets of M4 hardware, as shown in Figure 6-65.
3.) Refer to Figure 6-65. Attach three tie-wrap mounts, P/N 46-208747P2, to the left rail flange
with the remaining three of the sets of M4 hardware.
Important: Orient the tie-wrap mounts as shown in Figure 6-65. The tie-wraps should fasten
vertically through the adhesive mount.
Base CMD Adapter
Cable, P/N 2338086,
to J2
Part Number Description Quantity
2320158 Flange Ride 2
46-328417P4 M4 Socket-Head Cap Screw 16 (Eight per side)
46-328432P2 M4 Lock Washer 16 (Eight per side)
46-328430P2 M4 Flat Washer 16 (Eight per side)